Determination of oxygen r1 at 3 Tesla using samples with a concentration range of dissolved oxygen

Vatnehol, Svein Are Sirirud; Hol, Per Kristian; Bjørnerud, Atle; Amiry-Moghaddam, Mahmood; Haglerød, Camilla Margrete; Storås, Tryggve
Journal article; AcceptedVersion; Peer reviewed
View/Open
Determination+o ... ith+a+concentration....pdf (5.523Mb)
Year
2020
Permanent link
http://urn.nb.no/URN:NBN:no-79921

CRIStin
1760230

Metadata
Show metadata
Appears in the following Collection
  • Institutt for klinisk medisin [5635]
  • Fysisk institutt [2360]
  • Institutt for medisinske basalfag [1415]
  • CRIStin høstingsarkiv [15978]
Original version
Magnetic Resonance Materials in Physics, Biology and Medicine. 2020, 33 (3), 447-453, DOI: https://doi.org/10.1007/s10334-019-00783-x
Abstract
Objective

To investigate the sensitivity of modified Look–Locker inversion recovery (MOLLI) to measure changes in dissolved oxygen (DO) concentrations in water samples and to calculate sequence-specific relaxivity (r1m) and limit of detection (LOD).

Materials and methods

Ten water samples with a range of DO concentrations were scanned at 3 T using two variations of MOLLI schemes. Using linear regression the r1 of DO was estimated from the measured DO concentrations and T1 relaxation rates (R1). The results were combined with previously reported values on in vivo stability measures of the MOLLI sequences and used to estimate a LOD.

Results

DO concentrations ranged from 0.5 to 21.6 mg L−1. A linear correlation between DO and R1 was obtained with both MOLLI sequences, with an average correlation coefficient (R2) 0.9 and an average estimated r1 (r^1) of 4.45 × 10−3 s−1 mg−1 L. Estimated LOD was ≈ 10 mg L−1.

Conclusion

MOLLI T1-mapping sequences may be used for detecting dissolved oxygen in vivo at 3 T with an r^1 in the range 4.18–4.8 × 10−3 s−1 mg−1 L and a corresponding LOD for dissolved oxygen of approximately 10 mg L−1. MOLLI-based T1 mapping may be a useful non-invasive tool for quantification of in vivo changes of DO concentration during oxygen challenges.
 
Responsible for this website 
University of Oslo Library


Contact Us 
duo-hjelp@ub.uio.no


Privacy policy
 

 

For students / employeesSubmit master thesisAccess to restricted material

Browse

All of DUOCommunities & CollectionsBy Issue DateAuthorsTitlesThis CollectionBy Issue DateAuthorsTitles

For library staff

Login
RSS Feeds
 
Responsible for this website 
University of Oslo Library


Contact Us 
duo-hjelp@ub.uio.no


Privacy policy